Yokogawa’s new TDLS8000 houses all of the industry’s leading features in one robust device. The platform design is for in situ measurements which negate the need for sample extraction and conditioning. | Yokogawa Electric Corporation
| Measurement object | O2, CO, CO or CH4, CO2, CO + CO2, H2O, NH3, NH3+ H2O, H2S, HCl concentration in combustion exhaust gas and process gas |
|---|---|
| Measured components and range | Measured component |
| O2 | 0-1% |
| CO(ppm) | 0-200 ppm |
| CO+CH4 | CO |
| CH4 | 0-5 % |
| NH3 | 0-30 ppm |
| H2O (ppm) in non HC | 0-30 ppm |
| H2O (ppm) in HC | 0-30 ppm |
| CO (%) | 0-20 % |
| CO (%) + CO2(%) | 0-30 % |
| NH3+ H2O | NH3 |
| H2O | 0-5 % |
| H2S | 0-5 % |
| CO2(%) High Range | 0-1 % |
| CO2(%) Extend.Range | 0-30 % |
| H2O (%) | 0-10 % |
| HCl | 0-50 ppm |
| Hazardous area classifications | Division 1, Zone 1; Explosion-proof/ Flame-proof type; FM (US, Canada), ATEX, IECEx, Korea, NEPSI, Japan, EAC, INMETRODivision 2, Zone 2; Non-Incendive/Type n; FM (US, Canada), ATEX, IECEx, Korea, NEPSI, EAC, INMETRO, Japan |
| SIL Certification | The TDLS8000 expect digital output (2 points), digital input (2 points), valve control output (2 points), and digital communications (HART, Modbus/TCP) are certified in compliance with the following standard. IEC 61508:Functional safety of Electrical/electronic/programmable electronic related systems; SIL 2 capability for single analyzer use, SIL 3 capability for dual analyzer use. |
| Marine certificate | DNV Type Approval |
| Certification for Emission Monitoring | QAL1 |
| Display | Touchscreen 7.5 inch TFT color LCD panel, 640 x 480 (VGA) |
| Hazardous areaclassifications | Division 2, Zone2: Non-Incendive/Type n; FM (US, Canada), ATEX, IECEx, Korea, NEPSI, Japan |