Парамагнитный анализатор MG8E измеряет концентрацию кислорода в газовых смесях, в том числе содержащих горючие компоненты. Прибор выполнен во взрывозащищённом исполнении.
| Measurement object | Oxygen concentration in gaseous mixture |
|---|---|
| Measurement system | Paramagnetic system |
| Measurement range | 0-1 to 0-25 vol%O2, 3 ranges can be programmed. |
| Analog output signal | 4 to 20 mA DC (load resistance: Max. 550 Ω) |
| Contact output | Fail; 1 point, open or closed when error occurs, user configurable.Low auxiliary gas pressure alarm; 1 point, open or closed when error occurs, user configurable.Maintenance status; 1 point, closed during maintenance.Range Answerback or Hi/Lo Alarm; 2 points, normally de-energized. |
| Contact input | Remote range switching, calibration start |
| Sample Gas Condition | Gas flow; 300 to 800 ml/min. ±10% of a set value (Standard 600 ml/min)Gas pressure; Approx. 7 kPa in Analyzer inletGas temperature; 0 to 50℃ |
| Auxiliary Gas | N2gas, Pressure; 350 to 500 kPa |
| Calibration Method | Automatic calibration, Semiautomatic calibration, Manual calibration |
| Calibration gas | Zero gas; N2gasSpan gas; Dry air |
| Auxiliary gas pressure | N2gas, 350 to 500 kPa (Approx. 35 ml/min) |
| Ambient temperature | -5 to 50℃ |
| Installation | Indoor, panel or wall mounting |
| Structure | Flameproof (Ex d IIBT4X) |
| Power supply | 90 to 127 V AC, 48 to 63 Hz, 50 or 60 Hz |
| Repeatability | ±1% or less of span |
| Linearity | ±1% or less of span |
| Response time | 90% response within 3 sec; measured by analog output signal change after gas is through the analyzer inlet. |