Анализаторы серии FLEXA используются для непрерывных онлайн-измерений в промышленных установках, предлагая возможность измерения с одним или двумя датчиками, что делает их наиболее гибкими двупроводными анализаторами.
| Area Classification | General PurposeClass I Div II (without the use of IS Barrier)Class I Div I (with the use of IS Barrier) |
|---|---|
| Enclosure Material | Aluminum alloy die cast with epoxy coating (Standard offeringAluminum alloy die cast with urethane or high anti-corrosion coating (available upon request) |
| Housing Rating | IP66 (except Canada), Type 4X (except Canada), Type 3S/4X (Canada) |
| Power Supply | 2-wire 24VDC Loop Powered |
| Output Signal | Bi-directional HART digital communication, superimposed on mA (4-20mA) signal |
| Intrinsically safe type | ATEX, IECEx, FM, CSA and NEPSI, and nonincendive of FM and CSA |
| Input Range | Conductivity |
| Performance(The specifications are expressed with simulated inputs.) | Conductivity |
| ORP | Linearity: ±1 mV, Repeatability: ±1 mV, Accuracy: ±1 mV |
| Temperature | Linearity: ±0.3 ºCRepeatability: ±0.1 ºCAccuracy: ±0.3 ºC |
| Input Specification | The FLXA202 accepts output from membrane covered Dissolved Oxygen sensors. These sensors can be Galvanic type, where the sensor generates its own driving voltage or Polarographic type, where the sensor uses external driving voltage from the converter.The input range is 0 to 50 µA for Galvanic sensors and 0 to 1 µA for Polarographic sensors.For temperature compensation, the FLXA202 accepts Pt1000 (DO30 sensor) and NTC22k elements (OXYFERM and OXYGOLD sensors). |
| Display range | Conductivity |
| Resistivity | 0.005kΩ / K cm-1to 0.5MΩ /K cm-1: Accuracy: ±0.5%F.S.0.5MΩ / K cm-1to 1MΩ /K cm-1: Accuracy: ±1%F.S. |
| Performance(Accuracy)(The specifications are expressed with simulated inputs.) | Conductivity |
| Temperature compensation | NaCl table: ±1 %Matrix: ±3 % |
| Cable length | max. 60 meters total length of fixed sensor cable + WF10(J) extension cable.Influence of cable can be adjusted by doing an AIR CAL with the cable connected to a dry cell. |
| Step response | 90 % (< 2 decades) in 8 seconds |
| Measurement Range | DO Sensors |
| DO30G | Measurement range: 0 to 20 mg/l (ppm)Temperature: 0 to 40 ºC |
| Performance(Accuracy) (Performance in ppm mode) | Linearity |
| Repeatability | ±0.05 ppm or ±0.8% F.S., whichever is greater |
| Accuracy | ±0.05 ppm or ±0.8% F.S., whichever is greater |
Заполните форму — ответим с актуальной ценой и сроком поставки в течение рабочего дня